*May not be identical to image*
A universal reader device that can capture and interpret the results from traxSensors (or others) providing real-time data analysis and reporting.
traxReader V6.2
Gate/WE Voltage Range
-1500 mV to 1500mV (When Source is 1500mV)
Gate/WE Voltage Resolution
~45 µV
Gate/WE Voltage Stability
+/- 0.1 %
Gate/WE Voltage Accuracy
+/- 0.1 %
Gate/WE Response Time
~5 µs
Gate/WE Output Resistance
~2 Ω
Max Gate/WE Current
~60.7 µA
Gate/WE Current Resolution
~7.2 pA
Source/CE Voltage
0 to 3000mV (Gate & Drain are relative to Source)
Source/CE Voltage Stability
+/- 0.1 %
Source/CE Voltage Accuracy
+/- 0.1 %
Source/CE Response Time
~5 µs
Source/CE Output Resistance
~100 mΩ
Drain Voltage
-1500 mV to 1500mV (When Source is 1500mV)
Drain Voltage Stability
+/- 0.1 %
Drain Voltage Accuracy
+/- 0.1 %
Drain Response Time
~5 µs
Drain Output Resistance
~2Ω
Reference Voltage
0 to 3000mV
Reference Input Resistance
3 TΩ
Max Source/Drain Current Range 1
27mA
Source/Drain Current Resolution Range 1
3.6nA
Max Source/Drain Current Range 2
3mA
Source/Drain Current Resolution Range 2
0.36nA
Available Electrode Oversampling Rates
32 @ 38400Hz, 64 @ 19200Hz, 128 @ 9600Hz, 256 @ 4800Hz, 512 @ 2400Hz, 1024 @ 1200Hz, 2048 @ 600Hz, 4096 @ 300Hz, 8192 @ 150Hz , 16384 @ 75Hz, 20480 @ 60Hz, 24576 @ 50Hz, 40960 @ 30Hz, 49152 @ 25Hz, 81920 @ 15Hz, 98304 @ 12.5Hz